Ideal UPS protection for applications in medical patient-care areas where UL 60601-1 listed medical electrical equipment is required
Built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise while maintaining full AC surge suppression in all modes
Provides connected equipment with continuous power during short and long duration loss of commercial power
Expandable runtime is available with the addition of one or more external battery packs (BP36V15-2U limit 1 or BP36V42-3U multi-pack compatible)
Line-interactive voltage regulation corrects brownouts and overvoltages from 81 to 143 volts back to usable 120V output levels, without using battery power
Tripp Lite's SMART 1200XLHG offers medical-grade UPS protection with built-in line-interactive voltage regulation, expandable battery runtime and intelligent monitoring with unattended shutdown capability. UL 60601-1 listed as medical electrical equipment. Built-in isolation transformer with Faraday shield offers full line isolation and removes common mode noise while maintaining full AC surge suppression in all modes less than 100 microamps. Includes high quality hospital-grade plug and outlets. Resistant to electrostatic discharge. 1000VA power handling capacity supports computers, inter-networking equipment and other critical loads during commercial power failures. External battery connector allows configuration for up to several hours of runtime during power failures by adding one or more external battery packs (BP36V15-2U limit 1 or BP36V42-3U multi-pack compatible). Line interactive voltage regulation corrects brownouts as low as 81V and overvoltages as high as 143V back to normal 120V levels, saving battery reserves for use only during complete power failures. Includes diagnostic LEDs and 2 built-in communications ports (1 USB, 1 DB9) that work with included PowerAlert Software and cabling to provide automatic unattended shutdown of connected network and PC applications.
Ideal UPS protection for applications in medical patient-care areas where UL 60601-1 listed medical electrical equipment is required
Built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise while maintaining full AC surge suppression in all modes
Provides connected equipment with continuous power during short and long duration loss of commercial power
Expandable runtime is available with the addition of one or more external battery packs (BP36V15-2U limit 1 or BP36V42-3U multi-pack compatible)
Line-interactive voltage regulation corrects brownouts and overvoltages from 81 to 143 volts back to usable 120V output levels, without using battery power
Hospital-grade plug and outlets offer added degree of safety and reliability to input/output connections
Full line isolation prevents shock potential and operational problems in sensitive electronic equipment
Resistant to electrostatic discharge
Supports loads up to 1000VA/750W
28 min. half-load runtime (500VA/375W), 10 min. full load runtime (1000VA/750W)
4 protected hospital-grade outlets; 6-ft. cord
USB and DB9 communication ports
PowerAlert unattended shutdown software and complete cabling included
Complete surge protection, line isolation and EMI/RFI noise filtering
5 diagnostic LEDs indicate AC present, on battery, voltage correction, overload & battery low/replace
Battery replacement door and back panel grounding lug
Package includes SMART 1200XLHG Medical Grade Tower UPS System, PowerAlert Software and cabling, Warranty information, Instruction manual
Warranty - Yes
UL Rated - Yes
Spec Sheet available - Yes
Operating Temperature (Maximum):
104 F
Operating Temperature (Minimum):
32 F
Cord Length (in/mm):
70.87 in
Number of Outlets:
4
UL Rated:
Yes
AC Voltage:
115,120 Volts
Active Power:
750 Watt
Current Draw:
9.20 Amps
DC Voltage:
36 Volts
Power Factor:
0.80 %
Depth:
9.50 in
Height:
12.50 in
Width:
7.50 in
Product Weight:
45.20 lbs
Shipping Weight:
51 lbs
Warranty (Length/Type):
2 years
Rack Units (Height):
2U
Output Volt Amp Capacity (VA):
1000
Output kVA capacity (kVA):
1
Frequency Compatibility:
60 Hz
Output voltage regulation (line mode):
-18%, +8%
Output voltage regulation (Battery mode):
115V (+/- 5%)